| Good article. Imagine only having to buy one set of NOS tubes ever... haha...
REJUVENATION OF VACUUM TUBES
With the ever increasing scarcity of old
tubes, it is becoming more important that we try to save as many as
possible. Over the past two years I have been experimenting with the
rejuvenation of these older tubes. This work has been based primarily
on present day techniques used at Eimac and on data given in various
books published in the 1920's Using the methods described herein, I
have had approximately 85% success in returning inactive tubes back to
usable transconductance. The failures have primarily been due to
filaments being burned out during application of the excessive voltages
required. The tubes which failed either had filaments that had been
weakened from long hours of operation, or were marginal at the weld
joints. None of these failures were opened for investigation as they
are still valuable for display purposes.
The primary failure mode
of these older tube types is a loss of electron emission from the
filament or cathode. With the wide inter electrode spacing used in
these tubes, a short is very rare except in the case of a broken
filament wire where the oxide has flaked from the filament or cathode
and has touched the grid. The loss of electron emission typically shows
up in the tube tester as a weak tube or one which will not raise the
meter needle. If a tube tests normal and does not show any erratic
indication on the test meter, no attempt should be made to improve it
by rejuvenation.
The equipment required for rejuvenation is
relatively simple. In addition to a tube tester, a variable filament
supply is required with a meter of reasonable accuracy for measuring
the applied voltage. In place of a separate filament supply, a filament
voltmeter may be connected to the tube tester and the filament voltage
switch and "line" adjustment used for voltage control. For the
thoriated tungsten filaments it is preferable that no grid or plate
voltages be applied during rejuvenation. With the oxide emitter tube,
voltages should be applied during rejuvenation. The removal of plate
and grid voltages can be readily accomplished by the construction of an
adapter socket with filament connections only. The voltage applied to
the filament during rejuvenation mush be carefully controlled to the
values given. The accompanying graph shows the results of various
voltages applied to a thoriated tungsten filament during rejuvenation.
It shows that a voltage lower than the recommended value will
eventually result in a fairly good tube, while too high a voltage will
result in a tube which will remain weak.
Emission loss is
generally due to contamination (poisoning) of the emitting surface. The
vacuum and the original outgassing of the elements in these older tubes
was not near the present day standard, therefore, they contain
considerable residual gases. The poor emission usually is the result of
either the emitting surface being poor in storage, or, immediately upon
being heated the filament/cathode was poisoned by the residual gases
which had condensed on the emitting surface. The function of
rejuvenation is to drive off these condensed gasses and to replenish
the electron emitting layer on the surface of the filament/cathode.
Vacuum
tubes have essentially three basic types of emitters. These are: pure
tungsten, thoriated tungsten, or a directly or indirectly heated oxide.
The type of emitter in a given tube can be determined by its operating
color at rated filament voltage. The pure tungsten filament operated
bright white, the thorated tungsten filament runs orange to yellow,
while the oxide emitter operates in the dull red region.
The pure
tungsten filament needs little rejuvenation as its operating
temperature makes it self-cleaning. Operation at 110% of rated filament
voltage for up to 30 minutes should clean them up. This type of
filament was used in such tubes as the UV200, UV201, and in many types
of transmitting tubes.
The thorated tungsten filament is probably
the major one to be dealt with by the collector. This filament is a
composition of tungsten and thorium with the tungsten acting as the
heat source while the thorium is the emitting source. This filament was
used in tubes such as the UX200A, UX201A, UV99, UX99, UX120, UX210, and
in many of the later (and present day) transmitting tubes. Two methods
are used for rejuvenation of these filaments. If a tube is only weak or
gives erratic readings, the first procedure should be tried. If a tube
is completely dead (but the filament lights up) then the second
procedure should be used. 1) operate the filament at 135% of the rated
voltage for 30 minutes. Test the tube, and if the tube has improved but
is still not to rating, continue for another hour. If at the end of
this time the tube is still not up to specification, use the following
procedure. 2) In this procedure the filament is run white hot to strip
the emitting surface completely clean, then the surface is restored
using the above procedure. Operate the filament for 15 to 20 seconds at
350% of rated voltage with no other voltages applied. Then, operate the
tube under the conditions given in the first procedure. Test the tube
every 30 minutes, and if the tube is not up to rating after two hours,
it has reached the end of its useful life. Note: Do not attempt to test
the tubes at the end of the first step, as there will be no emission.
Typically
the oxide emitter consists of a layer of strontium and/or barium oxide
deposited on a heated surface. In the directly heated type, this layer
is placed directly on the surface of the filament. Typical of this type
are Western Electric tubes such as the VT-1 and VT-2 and the WD11,
UX226, and UX280. The indirectly heated cathode is the more modern type
of emitter consisting of a metal sleeve with the oxide layer on the
exterior and the filament mounted in the interior. The indirectly
heated cathodes include the ac heater types such as the 24, 27, and the
Kellog tubes. These tube types should initially be operated at the
rated filament voltage for at least one hour and then checked for
quality and stability. If they still are not satisfactory, then the
following procedure should be used. With the tube in the tube tester,
increase the filament voltage to 120% of rating while carefully
watching the plate current or tube tester meter reading. The meter
reading will slowly increase, hit a peak, then start to decrease. At
the point of maximum reading, reduce the filament voltage back to rated
value. Continue to operate the tube at rated filament voltage for at
least four hours, then test. When two tests spaced one hour apart
provide the same reading, the tube is rejuvenated as much as possible.
The
rejuvenation of the old tubes can be very rewarding especially
considering that some of the would otherwise be in the junk box. It
does take some time for this work as there are no short cuts, but it is
something that can be done without constant attendance. While not all
the tubes will come up to 100% or rating, at least many tubes can be
brought up to the point of being usable. As these old tubes become more
scarce this may be the only way we will have of getting the old sets
operating.
|